I)
2.The test bench control computer is sending the control command STBY (Stand- by) to the system unit or any single analyzer.
There are two different cases:
AIf the system or the analyzer is resting, this mode will be finished. Then, it will be tried to get the
Example:
The system or the analyzer is resting. No error is existing. The test bench control computer will ask for the operation mode:
Test bench control computer sends ASTZ Kn
→System or analyzer will response ASTZ 0 SREM SPAU
Test bench control computer sends STBY. System or analyzer shall accept the standby mode:
Test bench control computer sends STBY Kn
→System or analyzer will response STBY 0
The system or analyzer is finishing the resting. Then, it will try to get the
Status is error free:
Test bench control computer sends ASTZ Kn
→System or analyzer will response ASTZ 0 SREM STBY or status has still some errors:
Test bench control computer sends ASTZ Kn
→System or analyzer will response ASTZ x SREM STBY
Test bench control computer sends ASTF Kn
→System or analyzer will response ASTF x n
The error status byte will be zero and the system or the analyzer will be ready to measure, if all temperatures are in the allowed setpoint range, if the FID flame is burning etc.
If these parameters are not correct, the error status byte will be different from zero. The test bench control computer will read the operation mode and the error status as long as the system or the analyzer will be ready to measure. The test bench control computer will control the maximum time for this reading.
90003752(1) | AK | 1 - 19 |