II) V24/RS232/485 Interface - Single Analyzers and Systems
2. List of all Codes |
| |||
2.1. Control commands |
| |||
CODE |
|
| Function | Page |
SALI | Kn | Mm | Linearization check with spangas | |
SARA | Kn |
| Autoranging "OFF" (located range will remain) | |
SARE | Kn |
| Autoranging "ON" | |
SATK | Kn |
| Autom. calibration | |
|
|
| (zero and span calibration + zerogas + |
|
SEGA | Kn |
| Spangas (spangas will flow with time limit) | |
SEMB | Kn | Mm | Set range (1, 2, 3, 4) | |
SFRZ | Kn |
| Decimal point setup for floating point format | |
SGTS | Kn |
| Device test | |
SHDA | K0 |
| Hold mode "OFF" | |
SHDE | K0 |
| Hold mode "ON" | |
SINT | Kn |
| Start integration (integral average) | |
SLCH | Kn | Mm | Linearization check (with gas distribution) | |
SLIN | Kn | Mm | Linearization | |
|
|
| (with determination of corrections and saving) |
|
SLST | Kn |
| Set linearization step and get values | |
SMAN | Kn |
| Operation mode "Manual" | |
SMGA | Kn |
| Samplegas (will be sucked or will be on) | |
SNAB | Kn |
| Zerogas calibration | |
SNGA | Kn |
| Zerogas (zerogas will flow with time limit) | |
SPAB | Kn |
| Spangas calibration | |
SPAU | Kn |
| Pause (resting status, e.g. pumps, ozonator, deozonator, high | |
|
|
| voltage off, gas input closed within the device) |
|
SQEF | Kn | Mm | Cross interference (only for CO analyzers) | |
SREM | Kn |
| Operation mode "Remote" | |
SRES | Kn |
| Reset (analyzer will change via initializing mode to | |
SROF | Kn |
| Delay modus "OFF" (operation mode: signal output with/without | |
|
|
| delay time, involved are SINT, AKON, AIKO, AIKG, analog signal) |
|
SRON | Kn |
| Delay modus "ON" (analog to "SROF") | |
SSPL | Kn |
| Purging (purge air will be sucked or will be on) | |
ST9O | Kn |
| Set t90 time step (S = fast, M = medium, L = slowly) | |
STBY | Kn |
| ||
|
|
| history) |
|
Optional: |
|
|
|
|
SCAL | Kn |
| Start system calibration (only for platform) | |
SENO | Kn |
| NO measurement (operation mode CLD) | |
SNOX | Kn |
| NOx measurement (operation mode CLD) | |
Kn: | Channel n |
|
| |
Mm: | Range m |
|
| |
Response
The response to the control commands will contain the CODE of the control command and the error status byte
90003752(1) | AK | 2 - 5 |