II) V24/RS232/485 Interface - Single Analyzers and Systems

Stability control procedure

 

 

Wait dead time Tt, if

 

Parameters:

 

 

 

 

 

 

 

 

 

 

gases were switched

 

To

Time-out [xx s] (max. time for stability control)

 

 

 

 

 

 

 

 

Tt

Dead time [xx s]

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

(Wait after gas switching)

 

 

Start timer for time-out To

 

Ti

Integration time [xx,x s] (for K1, Knew)

 

 

 

Ts

Stability time [xx s]

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Tol.

Tolerance [x,x % range]

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Calculate concentration

 

 

 

 

 

 

 

 

mean value K1 over time Ti

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Set and start timer

 

 

 

 

 

 

 

 

for stability time

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Calculate concentration

 

 

 

 

 

 

 

mean value Knew over time Ti

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Deviation = K1 - Knew

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Time out ?

Yes

 

Signal not

 

 

 

 

 

 

 

stable

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

No

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Yes

 

 

 

 

 

 

 

 

Dev. Tol.

 

K1 = Knew

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

No

 

 

 

 

 

 

 

 

 

 

 

 

 

 

No

Stability

 

 

 

 

 

time over ?

Yes

Signal stable

Signal = Knew