SATK – Control command "Automatic calibration"

Starting this command the analyzers in a system or the single analyzer will start a calibration procedure to determine the correction values. The required calibration gases and the pumps will be switched on automatically. Then, the calibration procedure will start automatically. Such a procedure has to run until the end. It may not be canceled or interrupted by other functions. Otherwise it is not sure that the correction values will be valid to calculate the exhaust values of analysis. Exceptions are the commands "Reset" or "Stand-by". After the procedure will be over the system, the analyzers in a system or the single analyzer will have to change to the operation mode "Stand-by".

Control command

SATK K0

Function associated to the whole system unit

Code

SATK K1 Kn

Function associated to channel n

Function associated to channel 1

Code

SATK Kn M3

Function associated to channel n and range 3

Code

Response

SATK 0

Error status

Code

2 - 10

AK