2.2. Read commands |
|
| ||
CODE |
|
| Function | Page |
AAEG | Kn |
| Spangas deviation | |
AALI | Kn | Mm | Deviations of the last linearization checks with spangas | |
AANG | Kn |
| Zerogas deviation | |
AEMB | Kn |
| Selected range (R1, R2, R3, R4) | |
AFDA | Kn | CODE | ||
AFDA | Kn | CODE | Purge time | |
AGID | Kn |
| Device identification | |
AGRW | Kn | m | Limit | |
AIKG | Kn |
| Concentration integral average; all (in ppm) | |
AIKO | Kn |
| Concentration integral average (in ppm) | |
AKAK | Kn | Mm | Calibration gas concentration (in ppm) | |
AKAL | Kn |
| Stored calibration corrections | |
AKFG | Kn |
| Configuration of the system | |
AKON | Kn |
| Concentration (current value in ppm) | |
ALCH | Kn | Mm | Deviations of the last linearization checks | |
ALIK | Kn |
| Calculation of linearization curve | |
ALIN | Kn | Mm | Linearization values in the device (X/Y = | |
ALKO | Kn | Mm | Polynomial coefficents of the linearization curve | |
ALST | Kn |
| Linearization steps | |
AM90 | Kn |
| Actual response time (t90 time) | |
AMBA | Kn | Mm | Begin of range (in ppm) | |
AMBE | Kn | Mm | End of range (in ppm) | |
AMBU | Kn |
| Switch levels for autoranging | |
AMDR | Kn |
| Manual adjusted pressure | |
AQEF | Kn |
| Cross interference check result (in ppm) (only for CO analyzers) | |
ASOL | Kn |
| Setpoint value with limits | |
ASTA | Kn |
| General status of the system | |
ASTF | Kn |
| Internal error status | |
ASTZ | Kn |
| Device status | |
ASYZ | Kn |
| System time (year, month, day, hour, min., sec.) | |
AT9O | Kn |
| T90 time (response time) | |
ATEM | Kn | m | Temperature | |
ATOL | Kn | Mm | Stability tolerances | |
AVEZ | Kn |
| Delay and synchronization time | |
AZEI | Kn | CODE | Times (for procedures) | |
Optional: |
|
|
|
|
ABST | Kn |
| Counter of operating hours | |
ADRU | Kn | m | Pressure (for service) | |
ADUF | Kn | m | Flow (for service) | |
AKEN | Kn |
| Device tag | |
AKOW | Kn | Mm | Correction (zerogas calibration and gradient) | |
AUKA | Kn |
| Uncorrected analog value | |
Kn: | Channel n |
|
| |
Mm: | Range m |
|
| |
Response
The response to the read command will contain the CODE of the read command, the error status byte
2 - 6 | AK | 90003752(1) |