BThe system or the analyzer is in the operation mode SXYZ. This mode will be finished. Then, it will be tried to get the
Example:
The system or the analyzer is in the operation mode SXYZ. An error is existing with the error number n, i.e. FID flame is not burning. The test bench control computer will ask for the error status:
Test bench control computer sends ASTF Kn
→System or analyzer will response ASTF x n
Test bench control computer sends STBY. System or analyzer shall accept the standby mode and get ready for an error free measurement:
Test bench control computer sends STBY Kn
→System or analyzer will response STBY x
The system or analyzer is finishing the operation mode SXYZ. Then, it will try to get the
Error is removed (e.g. Flame was ignited):
Test bench control computer sends ASTZ Kn
→System or analyzer will response ASTZ 0 SREM STBY
or
Error is still existing (e.g. Flame has not been ignited):
Test bench control computer sends ASTZ Kn
→System or analyzer will response ASTZ x SREM STBY
Test bench control computer sends ASTF Kn
→System or analyzer will response ASTF x n
The error status byte will be zero and the system or the analyzer will be ready to measure, if the error is removed i.e. the FID flame is still burning.
If these parameters are not correct, the error status byte will be different from zero. The test bench control computer will read the operation mode and the error status as long as the system or the analyzer will be ready to measure. The test bench control computer will control the maximum time for this reading.