II) V24/RS232/485 Interface - Single Analyzers and Systems
SLIN – Control command "Linearization"
Starting this command the analyzers in a system or the single analyzer will switch on the gas tubing to a gas distribution. A linearization procedure for the selected range will run. The device will record the determined correction values to the receiver specific raw curve. The values will be stored in the device to calculate the gas concentration. This procedure will be controlled by several commands of the TBCC or the system. The device or the gas distribution will only accept those commands, it they have already received the "SLIN" command. The "SLIN" command prepares the device or the gas distribution to receive and execute further commands being necessary for the linearization procedure. The spangas concentration has to be set up before by the "EKAK" command.
Control command
SLIN Kn Mn
Range n
Function associated to channel n
Code
Response
SLIN 0
Error status
Code
90003752(1) | AK | 2 - 23 |