II) V24/RS232/485 Interface - Single Analyzers and Systems

SMAN – Control command "Operation mode MANUAL"

With this command the FU will change to the operation mode "Manual". Then it will only be possible to start functions from an operating unit integrated in the FU. The same operation mode will be enabled, if the service switch of the FU will be in the position "Remote Disable". In that mode it will only be possible to answer to read commands of the TBCC.

Control command

SMAN K0

Function associated to the whole system unit

Code

SMAN K1 Kn

Function associated to channel n

Function associated to channel 1

Code

Response

SMAN 0

Error status

Code

90003752(1) [AK-Commands] 10/98

AK