SPECIFICATIONS - General

Enclosure: Polycarbonate. NEMA 4X/CSA 4 (IP65).

Dimensions: Overall 155 x 155 x 131mm (6.10 x 6.10

x 5.15 in.). Cutout: 1/2 DIN 139mm x 139mm (5.45 x 5.45 in.)

Conduit Openings: Accepts 1/2” or PG13.5 conduit fittings

Display: Monochromatic graphic liquid crystal display. 128 x 96 pixel display resolution. Backlit. Active display area: 58 x 78mm (2.3 x 3.0 in.).

Ambient Temperature and Humidity: 0 to 55°C (32 to 131°F). Turbidity only: 0 to 50°C (32 to 122°F), RH 5 to 95% (non-condensing)

Storage Temperature Effect: -20 to 60ºC (-4 to 140°F)

Hazardous Location Approvals -

Class I, Division 2, Groups A, B, C, & D

Class Il, Division 2, Groups E, F, & G

Class Ill T4A Tamb= 50°C

Evaluated to the ANSI/UL Standards. The ‘C’ and ‘US’ indicators adjacent to the CSA Mark signify that the product has been evaluated to the applicable CSA and ANSI/UL Standards, for use in Canada and the U.S. respectively

Options for CSA: -01, 02, 03, 20, 21, 22, 24, 25, 26, 27, 30, 31, 32, 34, 35, 36, 37, 38, AN, and HT.

Note: Turbidity configurations (Models 1056-02-27-38- AN/-HT, 1056-03-27-38-AN/-HT, 1056-02-27-37-AN/ -HT and 1056-03-27-37-AN/-HT) are CSA approved Class I Div. 2 for hazardous area installation.

Class I, Division 2, Groups A, B, C, & D Class Il & lll, Division 2, Groups E, F, & G T4A Tamb= 50°C Enclosure Type 4X

Options for FM: -01, 02, 03, 20, 21, 22, 24, 25, 26, 27, 30, 31, 32, 34, 35, 36, 37, 38, AN, and HT.

Note: Turbidity configurations (Models 1056-02-27-38- AN/-HT, 1056-03-27-38-AN/-HT, 1056-02-27-37-AN/- HT and 1056-03-27-37-AN/-HT) are FM approved Class I Div. 2 for hazardous area installation.

Ordinary Locations-

The following ordering options are approved by UL: -01, 02, 03, 20, 21, 22, 23, 24, 25, 26, 27, 30, 31, 32, 33, 34, 35, 36, 37, 38, AN, DP, and HT.

Note: -UL order code must be specified for UL approval. Model 1056 UL configurations are assembled per UL requirements and are marked UL on the unit.

POLLUTION DEGREE 2: Normally only non-conductive pollution occurs. Occasionally, however, a temporary conductivity caused by condensation must be expected.

Altitude: for use up to 2000 meter (6562 ft.)

Power: Code -01: 115 VAC ±15% 60 Hz ±6%, 10W; 230 VAC ±15% 50 Hz ±6%, 10 W.

Code -02: 20 to 30 VDC. 15 W.

Code -03: 84 to 265 VAC, 47 to 63.0 Hz. 15 W.

Note: Code -02 and -03 power supplies include four programmable relays

Equipment protected by double insulation

RFI/EMI: EN-61326

LVD: EN-61010-1

Alarms relays*: Four alarm relays for process meas- urement(s) or temperature. Any relay can be config- ured as a fault alarm instead of a process alarm. Each relay can be configured independently and each can be programmed with interval timer settings.

Relays: Form C, SPDT, epoxy sealed

Maximum Relay Current

 

Resistive

 

 

28 VDC

5.0 A

115 VAC

5.0 A

 

 

230 VAC

5.0 A

 

 

Inductive load: 1/8 HP motor (max.), 40 VAC

Maximum screw torque for power lead connector and relay terminal blocks is 0.6 N.m

CAUTION

RISK OF ELECTRICAL SHOCK

*Relays only available with -02 power supply (20 - 30 VDC) or -03 switching power supply (84 - 265 VAC)

WARNING

WARNING

Exposure to some chemicals may degrade the sealing properties used in the following devices: Zettler Relays (K1-K4) PN AZ8-1CH-12DSEA

Inputs: One or two isolated sensor inputs

Outputs: Two 4-20 mA or 0-20 mA isolated current out- puts. Fully scalable. Max Load: 550 Ohm. Output 1 has superimposed HART signal (configurations 1056-0X-2X-3X-HT only)

Current Output Accuracy: ±0.05 mA @ 25 ºC

Terminal Connections Rating: Power connector (3-leads):18-12 AWG wire size. Signal board ter- minal blocks: 26-16 AWG wire size. Current output connectors (2-leads):24-16 AWG wire size. Alarm relay terminal blocks: 18-16 AWG wire size (-02 24 VDC power supply and -0384-265VAC power supply)

Weight/Shipping Weight: (rounded up to nearest lb or nearest 0.5 kg): 3 lbs/4 lbs (1.5 kg/2.0 kg)

5

Page 5
Image 5
Emerson Process Management 1056 instruction sheet Ordinary Locations, RFI/EMI EN-61326, Maximum Relay Current Resistive

1056 specifications

Emerson Process Management's 1056 is a high-performance flue gas analyzer designed for industrial applications requiring precise measurement of combustion and emissions. This advanced analyzer is instrumental in optimizing combustion processes, ensuring regulatory compliance, and enhancing overall operational efficiency.

One of the key features of the Emerson 1056 is its multi-gas measurement capability. The device can simultaneously monitor multiple gases, such as oxygen, carbon dioxide, carbon monoxide, and nitric oxides. This comprehensive measurement range provides operators with critical insights necessary for managing combustion processes effectively.

The 1056 utilizes advanced infrared and electrochemical sensing technologies. The infrared sensors are particularly effective for measuring gases like CO₂ and CO, while electrochemical cells handle specific gases like O₂ and NOx. These technologies offer high accuracy and reliability across a wide temperature and pressure range, ensuring that the data collected is both timely and precise.

Another significant characteristic of the Emerson 1056 is its robust construction, which is designed to withstand harsh industrial environments. The analyzer features durable materials that resist corrosion and high temperatures, ensuring long-lasting performance. Additionally, the 1056 is equipped with a user-friendly interface that simplifies operation and data interaction. Its intuitive design allows users to access real-time data and easily configure settings, facilitating quick decision-making on the shop floor.

For communication and integration with existing systems, the 1056 supports various industry-standard protocols, such as HART, Modbus, and Ethernet/IP. This compatibility enables seamless integration into broader control and monitoring systems, allowing for enhanced data sharing and management.

Emerson has also placed a strong emphasis on serviceability with the 1056. The analyzer is designed for easy maintenance, with features that simplify calibration and routine checks. This reduces downtime and operational disruptions, ensuring continuous compliance with environmental regulations.

In conclusion, the Emerson Process Management 1056 stands out as a versatile and reliable flue gas analyzer, offering advanced measurement capabilities, robust construction, and seamless integration into industrial systems. Its combination of technology and user-centric design makes it an essential tool for industries striving for efficiency and compliance in their combustion processes.