The parameter is indicated by,
CtorSt Cut-out reset mode parameter
Press “SET” to access the parameter setting. Normally the cut-out is set for au-
tomatic mode.
Cto=Auto Cut-out set for automatic reset
Tochange to manual reset mode press “UP” and then “SET”.
Cto=rSt Cut-out set for manual reset
8.11.2.3 Soak Stability
The soak stability controls the required stability of the well temperature for the
soak time (Section 8.6.3). The stability is in degrees Celsius. The default is
0.1°C. This valuecan be changed in the parameter menu.
8.11.3 Serial Interface Parameters
The serialRS-232 interface parameters menu is indicated by,
SErIAL Serial RS-232 interface parameters menu
The serialinterface parameters menu contains parameters which determine the
operation of the serialinterface. These controls only apply to instruments fitted
with the serial interface.The p arameters in themenu are — BAUDrate, sample
period, duplexmode, and linefeed.
8.11.3.1 BAUD Rate
The BAUDrate is the first parameter in the menu. The BAUD rate setting de-
termines the serial communications transmission rate.
The BAUDrate parameter is indicated by,
BAUd Serial BAUD rate parameter
Press “SET” to choose to set the BAUDrate. The current BAUD rate value will
then be displayed.
1200 b Current BAUD rate
The BAUDrate of the serial communications may be programmed to 300, 600,
1200, or 2400 BAUD.Use “UP” or “DOWN” to change the BAUD rate value.
2400 b New BAUD rate
9105/9107 39
8Controller Operation