Maintenance, Calibration, and Troubleshooting
Maintenance 3
Section 3
Maintenance, Calibration and Troubleshooting
3.1 Maintenance
No periodic maintenance is required for the module. The output voltage levels, however, should be verified periodically.
NOTE
If a maintenance question arises and cannot be resolved by using this manual, please contact Fluke Biomedical at 440.248.9300.
3.2 Calibration
There are no user adjustments on the main
Adjustments to the
3.3 Troubleshooting
WARNING
Extreme care must be used when troubleshooting a system that has power applied. All standard troubleshooting precautions apply.
WARNING
Once a problem has been located, remove all power before continuing with the repair.
CAUTION
Personnel performing the following procedure must be familiar with the operation of the monitoring system and the location of each piece of equipment used in the system.