Fluke 960PS-210, 960PS-200 manual Installation, Procedures, Warnings, and Cautions, Introduction

Models: 960PS-210 960PS-200

1 16
Download 16 pages 48.24 Kb
Page 7
Image 7

Introduction

1

Receiving Inspection and Storage

2.Requirements for proper storage must be documented and written procedures or instructions must be established.

3.In the event of fire, post-fire evaluation must be in accordance with ANSI N45.2.2 (1972), Section 6.4.3.

4.Removal of items from storage must be in accordance with ANSI N45.2.2 (1972), Sections 6.5 and 6.6.

1.5 Installation

The 960PS-200 and 960PS-210 Power Supply Modules are supplied as part of a radiation monitoring system (RMS) or as replacement parts for an existing RMS. When the power supplies are shipped as part of a system, they are installed at the factory.

When a module is shipped as a replacement part, verify the integrity of the gasket, heat sink, and seal screw fasteners (when used) prior to installation. Apply P/N 88-10 thermal coat, or equal heat transfer compound between power supply and heat sink to ensure the thermal coupling of the installed power supply. For operation of the 960PS-210 at 220 VAC, verify the location of the internal supply voltage jumper E1. The part number of the required mounting hardware may be found in the system assembly drawing section of the applicable systems manual.

1.6 Procedures, Warnings, and Cautions

The equipment described in this manual is intended to be used for the detection and measurement of ionizing radiation. It should be used only by persons who have been trained in the proper interpretation of its readings and the appropriate safety procedures to be followed in the presence of radiation.

Although the equipment described in this manual is designed and manufactured in compliance with all applicable safety standards, certain hazards are inherent in the use of electronic and radiometric equipment.

WARNINGS and CAUTIONS are presented throughout this document to alert the user to potentially hazardous situations. A WARNING is a precautionary message preceding an operation that has the potential to cause personal injury or death. A CAUTION is a precautionary message preceding an operation that has the potential to cause permanent damage to the equipment and/or loss of data. Failure to comply with WARNINGS and CAUTIONS is at the user's own risk and is sufficient cause to terminate the warranty agreement between Fluke Biomedical and the customer.

Adequate warnings are included in this manual and on the product itself to cover hazards that may be encountered in normal use and servicing of this equipment. No other procedures are warranted by Fluke Biomedical. It shall be the owner's or user's responsibility to see to it that the procedures described here are meticulously followed, and especially that WARNINGS and CAUTIONS are heeded. Failure on the part of the owner or user in any way to follow the prescribed procedures shall absolve Fluke Biomedical and its agents from any resulting liability.

Indicated battery and other operational tests must be performed prior to each use to assure that the instrument is functioning properly. If applicable, failure to conduct periodic performance tests in accordance with ANSI N323-1978 (R1983) Radiation Protection Instrumentation Test and Calibration, paragraphs 4.6 and 5.4, and to keep records thereof in accordance with paragraph 4.5 of the same standard, could result in erroneous readings or potential danger. ANSI N323-1978 becomes, by this reference, a part of this operating procedure.

1-3

Page 7
Image 7
Fluke 960PS-210 manual Installation, Procedures, Warnings, and Cautions, Introduction, Receiving Inspection and Storage