Overload protection for all functions and ranges: 600 V.

Function

Input Impedance (Nominal)

L, mL, K

L, mL

K

>10 MΩ, <50 pF

Common Mode

Rejection Ratio (1

kΩ Unbalanced)

>120 dB at dc, 50 Hz, or 60 Hz

>60 dB dc to 60 Hz

Normal Mode

Rejection

>60 dB at 50 Hz or 60 Hz

e

G R

Open Circuit Test Voltage

<3.1 V dc

<2.8 V dc (typical)

<3.1 V dc

Full Scale Voltage

To 3.2 MΩ

 

32 MΩ

 

 

 

<440 mV dc

 

<1.4 V dc

<420 mV dc

 

<1.3 V dc

(typical)

 

(typical)

 

 

 

2.0

V dc

 

 

 

e

GR G

Typical Short Circuit Current

 

400 μA

 

500 μA

 

 

 

 

VF

 

Typical Test Current

0.0 V

 

0.5 mA

0.6 V

 

0.4 mA

1.2 V

 

0.3 mA

2.0 V

 

0.1 mA

Page 12