8 - English [EN] K386 Issue 1

Specification data

All accuracy statements are for one year.

Specification - General

Specification - Temperature ranges

Languages English [Default]
Operating temperature 14 … 122°F (-10 … 50°C)
Storage temperature -4 … 140°F (-20 … 60°C)
Humidity 0 to 90% without condensation
(Def Stan 66-31, 8.6 cat III)
Shock/Vibration BS EN 61010:2001; Def Stan 66-31, 8.4 cat III
EMC BS EN 61326-1:1998 + A2:2001
Safety Electrical - BS EN 61010:2001; CE Marked
Size (L: W: H) 7.1 x 3.3 x 2.0 in (180 x 85 x 50 mm)
Weight 14 oz (400 g)
Power supply 3 x AA alkaline batteries
Duration > 80 hours (Measure),
> 10 hours (24 V Source at 12.5 mA)
Thermocouple type Standard Range °F Range °C Accuracy °F * Accuracy °C *
K IEC 584 -454 .... -328 -270 .... -200 3.6 2.0
K IEC 584 -328 ... 2498 -200 ... 1370 1.1 0.6
J IEC 584 -346 ... 2192 -210 ... 1200 0.9 0.5
T IEC 584 -454 .... -292 -270 .... -180 2.5 1.4
T IEC 584 -292 ...... -94 -180 ...... -70 0.9 0.5
T IEC 584 -94 ..... 752 -70 ..... 400 0.6 0.3
B IEC 584 122 ..... 932 50 ..... 500 7.2 4.0
B IEC 584 932 ... 2192 500 ... 1200 3.6 2.0
B IEC 584 2192 ... 3308 1200 ... 1820 1.8 1.0
R IEC 584 -58 ....... 32 -50 ......... 0 5.4 3.0
R IEC 584 32 ..... 572 0 ..... 300 3.6 2.0
R IEC 584 572 ... 3216 300 ... 1769 1.8 1.0
S IEC 584 -58 ....... 32 -50 ......... 0 4.5 2.5
S IEC 584 32 ..... 212 0 ..... 100 3.4 1.9
S IEC 584 212 ... 3216 100 ... 1769 2.5 1.4
E IEC 584 -454 .... -238 -270 .... -150 1.6 0.9
E IEC 584 -238 ... 1832 -150 ... 1000 0.7 0.4
N IEC 584 -454 ........ -4 -270 ...... -20 1.8 1.0
N IEC 584 -4 ... 2372 -20 ... 1300 1.1 0.6
L DIN 43710 -328 ... 1652 -200 ..... 900 0.6 0.3
U DIN 43710 -328 ..... 212 -200 ..... 100 0.9 0.5
U DIN 43710 212 ... 1112 100 ..... 600 0.6 0.3
C 32 ... 2732 0 ... 1500 1.8 1.0
C 2732 ... 3632 1500 ... 2000 2.5 1.4
C 3632 ... 4208 2000 ... 2320 3.4 1.9
D 32 ... 3092 0 ... 1700 1.8 1.0
D 3092 ... 3992 1700 ... 2200 2.9 1.6
D 3992 ... 4523 2200 ... 2495 6.5 3.6
*Cold Junction (CJ) error (Maximum):
Range 10 ... 30°C (50° ... 86°F) = 0.2°C (0.4°F)
Add 0.01° CJ error / ° ambient temperature change for ranges: 14 … 50°F, 86 … 122°F
(-10 … 10°C, 30 … 50°C)