4 - English [EN] K386 Issue 1
To start - Set up the settings
When the task is set up (Table2), use the Settings menu to
adjust the input operation.
If there is additional data for a menu option, select Settings(■■)
to see the values that are set up. If necessary, adjust the values.
Refer to “Edit functions”.
Table 3: Menu options - Settings
To start - Edit functions
Example 1) Set up a label for x:y Scaling = %.
Example 2) Set up values for x:y Scaling = 0 to 100%.
OperationThis section gives examples of how to connect and use the
instrument. Before you start:
• Read and understand the “Safety” section.
• Do not use a damaged instrument.
Operation - Thermocouple connections
Attach the thermocouple wires to the applicable thermocouple
mini-connector (Figure1). The wider blade is the negative.
Then attach the connector to the instrument.
Operation - Communications port connections
Use the communications port (A1- item [9]) to attach an IDOS
Universal Measurement Module (UMM) or an RS232 device.
• IDOS Communications
When you attach the cable from a UMM (Figure3), the
instrument automatically changes the menus to give you all the
applicable options (Table2).
• RS232 Communications
Use an RS232 cable (A1 - item [9a]) to transmit all the Snapshot
data to a PC or a printer (Table4).
Display: Task
T1, T2 1Settings selection
(Ifapplicable) 23
■ ■▲
▼
Sk2 = Settings
Menu:
Settings 456 Settings
x:y
▲
▼■ ■
(Table3) [✓]/[ ] Sk1 = Edit
Options Description
To include maximum, minimum and average values for
the measurement task (If applicable). Only available
when there is one input measurement task.
To select and set up a tare value for the measurement
task (a specified value or the reading on the display).
The instrument subtracts a positive tare value, and
adds a negative tare value.
Additional data: Select Settings (■ ■)
To select and set up a scale of values: One local scale
for each measurement task (Maximum: 5).
Additional data (Table1): Select Settings (■ ■)
To select and set up the filter values to give a
smoother output for the measurement task:
Band as a % of full scale (FS). The filter
compares each new value with the previous
value. If the new value is outside the band, it is
not filtered.
Low pass filter time constant in seconds.
Increase the value to increase damping factor.
Additional data: Select Settings (■ ■)
To select and set up the alarm values for the
measurement task (maximum and minimum).
Additional data: Select Settings (■ ■)
UMM - Pressure options only. Gage sensors or
sensors with differential operation. A zero correction
that makes the instrument read zero at local pressure.
Leak Test only. To set an applicable perio d fo r the leak
test (Hours:Minutes:Seconds).
Settings
x:y 1Edit 2Edit 3
■ ■
▲
▼
■ ■
Sk1 = Edit A2 - item [22]
Edit ... Edit 45
%_
▲
▼
■ ■
A2 - item [22]
Settings
x:y 1Settings
x:y 2Edit 34
▲
▼
■ ■
▲
▼
■ ■
Sk1 = Edit
Shift
OK