4.3.1
There are two types of cleaners available, a low pH cleaner (Calcium/Iron removal) and a high pH cleaner (Silt/Organics removal). GE Osmonics provides these cleaners in liquid form. The low pH cleaner attacks mineral scale deposits such as calcium, magnesium and iron
CAUTION:
The different cleaners must not be mixed together. Introduce the low pH cleaner first, thoroughly rinse the system, and then use the high pH cleaner.
WARNING:
The solution is a mild irritant to eyes and skin. It is recommended that protective gloves and eye protection be worn when handling the cleaner. Neither respiratory protection nor special ventilation are required.
Materials Required:
LIQUICLEAN 103™ – (Low pH liquid) - GE Osmonics Part No. 1228653 (1 Gal) LIQUICLEAN 511™ – (High pH liquid) - GE Osmonics Part No. 1228791 (1 Gal)
•high and low pH cleaner
•2 Gallons pretreated water
•Disinfect container with volume of 2 ½ gallon or more
•pH test strips
•phosphate testing kit (optional)
•product rinse line (clear hose with male QD fitting on one end, open on the other)
•“DO NOT USE” warning labels (not provided by GE Osmonics)
•cleaning / disinfect log sheets
•most current log sheets for water quality comparison
•sterile gauze and tape
4.3.1.1
Read and understand instructions on cleaner container before beginning procedure.
1.Connect the water inlet line to the water supply and the waste line to drain. Turn the RO unit and water supply on.
2.Close the product valve. Disconnect the product line (white QD) from inlet line of the Fresenius dialysis machine. Cover exposed inlet water line with sterile gauze and secure with tape.
3.Connect the product rinse line to the product line and direct to clean drain receptacle. Open the product valve.
12552e – 27 Feb 04 | 4- 12 | Routine Maintenance |