6.Ensure that the water supply to the RO unit is on. If it is partially closed, rotate the dialyze/flush valve to the full flush position.
7.Turn ON the
CAUTION:
Some pump noise (chattering) is normal in the flush position. Excessive pump noise (chattering) can be damaging to the pump. The RO is in “flush” from the full open (flush) position to just before the fully closed (dialyze) position. Adjust it by partly closing the flush valve to reduce chattering of the pump.
8.Rotate the dialyze/flush valve to the dialyze position.
9.Turn the
A.During this rinse period, rotate the Dialyze/Flush valve to the Flush position briefly and then back to Dialyze three times to
10.Test the product water for
11.When all of the parameters (% Rejection, TDS, pH) have normalized, operate the RO for an additional 20 minutes in rinse mode.
NOTE:
Repeat the cleaning procedure with the other pH cleaner. GE Osmonics recommends using a low pH cleaner first, followed by a high pH cleaner after rinsing thoroughly.
12.Turn the
13.Close the product valve and then disconnect the clear accessory flush tubing from the product line (white QD / valve) and store properly. Reconnect the product line to its original connection and open the product valve.
14.Remove the “Do Not Use/Contains Cleaner” warning label.
CAUTION:
The
12552e – 27 Feb 04 | 4- 15 | Routine Maintenance |