4.3.2BIOSAN
BIOSAN is a liquid cleaner/descaler agent intended for use in removing mineral scale and organic biofilm and is a
WARNING:
The solution is a mild irritant to eyes and skin. It is recommended that gloves (latex/vinyl) and eye protection be worn when handling BIOSAN. Neither respiratory protection nor special ventilation are required.
CAUTION:
BIOSAN is not intended for use in Hemodialysis Machines.
NOTE:
BIOSAN can be used as a low pH cleaner or as a short term (two weeks) storage solution.
Materials Required:
BIOSAN - GE Osmonics Part No. 14962 (1 Gal)
•product water
•Disinfect container with volume of 2 ½ gallon or more
•pH test strips
•product rinse line (extension hose)
•“DO NOT USE” warning labels (not provided by GE Osmonics)
•cleaning / disinfect log sheets
•most current log sheets for water quality comparison
•sterile gauze and tape
4.3.2.1 BIOSAN Cleaning Procedure
Read and understand all BIOSAN cleaning instructions before beginning procedure.
1.Connect the water inlet line to the water supply, and the waste and product lines to drain. Turn the RO unit and water supply on.
2.Establish the
3.Turn the RO OFF.
4.Close the product valve. Disconnect the product line (white QD) from inlet line of the Fresenius dialysis machine. Cover exposed inlet water line with sterile gauze and secure with tape.
5.Connect the product rinse line to the product line and direct to clean drain receptacle. Open the product valve.
12552e – 27 Feb 04 | 4- 16 | Routine Maintenance |