CAUTION:

Check that there is waste flow within 10 seconds. If not, turn off the RO and verify that the disinfect-line is connected properly.

CAUTION:

Do not allow air or foam to be drawn into the RO system, as this will damage the pump.

NOTE:

The % rejection will fall below 85% and the TDS will rise considerably when the cleaner is re-circulated through the F-801 RO unit.

NOTE:

The cleaners will foam; this is normal. If using a closed container, provide a vent to the atmosphere to relieve pressure.

3.Verify that the F-801 and Fresenius dialysis unit are properly labeled with warning signs to prevent use during cleaning process.

WARNING:

The F-801 RO machine in now completely filled with cleaning solution and must NOT be used to supply water to the Fresenius dialysis machine or for any other purpose until a complete rinse procedure has been performed. See Cleaner Rinse Procedure below.

4.Let the RO stand dormant with cleaning solution in it for sixty minutes with a maximum of overnight.

4.3.1.3 Post-Cleaning Rinse Procedure

1.After the dwell time, remove the waste hose and clear accessory flush line from the cleaning solution and place it in the drain with the appropriate air gap.

2.Turn the RO on and run cleaning solution to the drain until the container is empty. This aids in emptying the container of cleaning solution.

CAUTION:

DO NOT ALLOW THE PUMP TO RUN DRY. As an alternative, the solution may be poured in the drain.

CAUTION:

Do not allow air or foam to be drawn into the RO system, as this will damage the pump.

3.When the disinfect/cleaning tank is empty, turn the F-801 RO unit OFF.

4.Reconnect the feed line to the pre-filter outlet (green QD).

5.Open the product valve.

12552e – 27 Feb 04

4- 14

Routine Maintenance