Millenium™ RO System by GE Osmonics
1.5Environmental Considerations
Prior to the installation of the Millenium™ RO System, it will be necessary to provide utilities and create an environment suitable for the trouble free operation of the RO system and its accessories.
POWER:
The Millenium™ operates on 115 VAC
WATER:
Reverse osmosis systems require a reliable water source. The Millenium™ RO needs a minimum of
DRAIN:
A drain outlet is required by the RO system. The drain must have a minimum capacity of 2 gallons per RO per minute of continuous flow for as long as and as often as the RO is operating.
TEMPERATURE:
The Millenium™ RO is normally connected to a cold water supply. The performance specifications of the RO are based on 77°F (25°C) feed temperature. Each degree Fahrenheit the feed water temperature falls, the RO product flow decreases approximately 1.5%, and each degree Celsius drop, the product flow is reduced by 3% (See Technote 113 for the “Temperature Correction Factors” chart). In climates where the winter water temperature drops significantly below 77°F it may be necessary to temper (heat) the water.
WARNING:
The temperature of the tempered water should be monitored closely with a thermometer as high temperatures can damage the RO membranes and possibly harm the patient. Refer to the inlet water requirements (Section 1.10.1) for the maximum operating temperature for the Millenium™ RO.
CAUTION:
It is necessary to keep the RO from freezing to prevent serious damage to the RO unit.
NOISE:
The RO unit does not generate significant noise levels. Hard walls may reflect noise and will make the RO seem louder.
14693e – 31 Mar 03 | 1 - 5 | General Information |