Millenium™ RO System by GE Osmonics
Product Water Pressure:
Product Water Pressure is read on a 0 to 100 PSI gauge on the front panel. Product pressures will register only when the product is back pressured (product hose removed from fitting or connected to dialysis unit).
Product Recirculation Check Valve:
Prevents
Product Water Sensor:
The Product Water Sensor is used in conjunction with the inlet water sensor by the
Product Check Valve:
The Product Check Valve is used to prevent anything from flowing backward into the product side of the RO membranes.
CAUTION:
Total Dissolved Solids (TDS) Sensor:
Samples the product water quality via the sensor probe, and the final quality of the product water is indicated on a digital display on the front panel as TDS in mg/L and ppm. The monitor is temperature compensated to adjust the display value to changing water temperature.
Disinfect Inlet Port:
The Disinfect Inlet Port is a
WARNING:
Water from this port is NOT RO product water and should not be used for dialysis solutions.
Disinfect Switch:
The Disinfect Switch is used to introduce disinfectants and cleaners into the RO. For safety, this switch must be held continuously to operate.
CAUTION:
When drawing disinfectants or cleaning solutions, do not allow the RO to draw air. This may cause irreparable damage to the RO pump.
14693e – 31 Mar 03 | 1 - 8 | General Information |