Millenium™ RO System by GE Osmonics

WATER PURITY:

Plumbing materials can significantly contribute to the contamination of the water. Metallic pipes (with the exception of 316 stainless steel) must be eliminated once the purification process has begun. Schedule 40 or 80 PVC pipes, polypropylene, PVDF, and other FDA recommended materials are suitable for most grades of purified water. Care must also be exercised in the choice of a thread sealant. Teflon tape is suitable for all threaded connections in this system.

WARNING:

Ordinary pipe dope must be avoided since it may leach objectionable and potentially dangerous impurities into the water.

1.6Theory of Operation

The Millenium™ Portable Reverse Osmosis system (RO) is a fully cabinetized unit that purifies a given feed water (by way of reverse osmosis) for use in hemodialysis applications. The Millenium™ RO possesses monitoring for pressure, product water quality and membrane performance (percent rejection). The RO pump provides the pressure required to push water through the RO membranes. The fluid controls provide a means of managing flow rates and pressures. The following paragraphs detail the function of these fluid control components, which are used within the RO.

14693e – 31 Mar 03

1 - 6

General Information