18 GOSSEN METRAWATT GMBH
7.4 Recording: Duration and Type – durA, CYCLE Recording without Time Limits – durA
Key Sequence: SEt ... ... durA on oFF
Settings: oFF = time limiting disabled
Recording is continued until the memory is full. If the CYCLE function has been activated,
memory is overwritten in a cyclical fashion. See also drawing in chapter 7.3, page 16.
Time-Limited Recording – durA
Key Sequence: SEt ... ... durA ON
dd hh:mm
Settings: d = number of days (adjustable from 0 to 6)
hh = number of hours, mm = number of minutes
Recording is ended after the predetermined duration has expired.
See also drawing in chapter 7.3, page 16.
Cyclical Recording – CYCLE
Key Sequence: SEt rAtE ... CYCLE OFF on
Settings: on = cyclical recording activated
When the memory is full, the oldest data are overwritten with new data (FIFO). See also
drawing in chapter 7.3, page 16.
Re cord Until Memory is Full – CYCLE
Key Sequence: SEt ... ... CYCLE on OFF
Settings: OFF = cyclical recording deactivated
When the memory full, data recording is ended. See also drawing in chapter 7.3, page 16.
7.5 Internal Time – timE Set Current Weekday and Time (no date)
Key Sequence: SEt ... ... tiME dd hh:mm
Settings: dd = current weekday
hh:mm = current time in hours and minutes
Current weekday and time are set with this function.
7.6 Select Address and Modem Option – Addr, ModEM
Key Sequence: SEt ... ... Addr nn ModEM no YES
Settings: nn = 1 ... 15
ModEM = YES, if the adapter is
connected to the PC with a modem
If several memory adapters are connected to the PC, a separate address must be assigned
to each. Address number 1 should be assigned to the first adapter, address number 2 to the
second and so forth.
Note: If only one adapter has been connected, it should be set to address number 1.