7 maIntenanCe and CleanIng

WARNING: Failure on the part of the responsible facility employing the use of this equipment to implement a satisfactory maintenance schedule may cause undue equipment failure and possible health hazard.

maIntenanCe

WARNING: Before use, inspect the Pocket Doppler and probe(s) to ensure there is no visible evidence of damage which may affect the functioning of the device or pregnant woman’s safety, or create the potential for the device to operate in an unsafe manner. Thereafter, inspect the device for evidence of damage at least once each week. If damage is evident or suspected, contact GF Health Products, Inc. Technical Service

at the number on the back cover or your GF Health Products, Inc. authorized distributor before use.

WARNING: The following safety checks/tests must be performed at least once every two years or

as specified in your facility's test and inspection protocol, by a qualified person with adequate training, knowledge, and practical experience:

Inspect the equipment for mechanical and functional damage.

Ensure the safety labels are legible.

Verify that the device functions properly as described in this manual.

Test according to the pregnant woman's leakage current (IEC 60601-1/1988: Limit: 100 uA (B)).

The leakage current should never exceed the specified limit. Record the data and store with the Grafco Pocket Doppler. If the device is not functioning properly or fails any of the above tests, contact GF Health Products, Inc. Technical Support at the phone number on the back cover, or your GF Health Products, Inc. authorized distributor.

14423-INS-LAB-RevE13

25