Graham Field 14423-2 Recommended separation distances, Rated Separation distance according to

Models: 14423-2

1 44
Download 44 pages 38 Kb
Page 38
Image 38

Info 1: At 80 MHz and 800 MHz, the higher frequency range applies.

Info 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures, objects and people.

a:Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land mobile radios, amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicted theoretically with accuracy. To assess the electromagnetic environment due to fixed RF transmitters, an electromagnetic site survey should be considered. If the measured field strength in the location in which the Grafco Ultrasound Pocket Doppler is used exceeds the applicable RF compliance level above, the Grafco Ultrasound Pocket Doppler should be observed to verify normal operation. If abnormal performance is observed, additional measures may be necessary, such as reorienting or relocating the Grafco Ultrasound Pocket Doppler.

b:Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3 V/m.

recommended separation distances

recommended separation distances between portable and mobile rf communications equipment and the grafco ultrasound pocket doppler

The Grafco Ultrasound Pocket Doppler is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled. The customer or the user of the device can help prevent electromagnetic interference by maintaining a minimum distance between portable and mobile RF communications equipment (transmitters) and the Grafco Ultrasound Pocket Doppler as recommended below, according to the maximum output power of the communications equipment.

rated

separation distance according to

maximum

frequency of transmitter (m)

output power

 

 

80 mhz to 800 mhz

800 mhz to 2.5 ghz

of transmitter

 

 

(W)

 

 

 

 

 

0.01

0.1167

0.2334

 

 

 

0.1

0.3689

0.7378

 

 

 

1

1.1667

2.3334

10

3.6893

7.3786

 

 

 

100

11.6667

23.3334

14423-INS-LAB-RevE1338

Page 38
Image 38
Graham Field 14423-2 Recommended separation distances, Rated Separation distance according to, Frequency of transmitter m