6 Troubleshooting Guide

SYMPTOM

CHECK / ACTION

 

 

Concentrator power is

Check if concentrator is unplugged; plug in.

on and alarm is on, but

 

concentrator doesn’t

Check if circuit breaker is tripped; fix circuit

work

breaker.

 

If concentrator still doesn’t work, contact

 

Graham-Field equipment provider.

SYMPTOM

CHECK / ACTION

 

 

Concentrator works,

Inspect tubes and cannula to ensure that they are

but little or no oxygen

not kinked or blocked. Smooth tubes to unblock.

from outlet

 

 

Check humidifier bottle, if so equipped, for leaks

 

or jamming. Give humidifier bottle a slight shake

 

to open safety valve. Clean humidifier bottle,

 

reinstall and tighten.

 

If condition persists, contact Graham-Field

 

equipment provider.

SYMPTOM

CHECK / ACTION

 

 

Flowmeter ball located

Adjust flowmeter knob. If condition persists,

on “0” position or float

contact Graham-Field equipment provider.

is unstable

 

SYMPTOM

CHECK / ACTION

 

 

Concentrator works,

If yellow light blinks, and alarm sounds at low

but makes abnormal

intervals, or yellow light illuminates continuously

sound

and alarm sounds, contact Graham-Field

 

equipment provider.

SYMPTOM

CHECK / ACTION

 

 

Concentrator works, but yellow fault light lluminates (JB0160-015B, JB0160-015, JB0160-015B-220, JB0160-015-220 only)

OK to continue use, but contact Graham-Field equipment provider immediately. Concentrator’s oxygen concentration is safe, but decreasing. When oxygen concentration is too low for safe use, concentrator will shut off and optional O2 sensor red warning light will illuminate.