Graham Field JB0160-010, JB0160-015B-220 Maintenance, Cleaning the Cabinet, Bottom filter

Models: JB0160-015 JB0160-015B JB0160-010B

1 28
Download 28 pages 8.54 Kb
Page 13
Image 13
5 Maintenance

5 Maintenance

Cleaning the Cabinet

Clean the cabinet at least once a month.

1.Turn off the power switch.

2.Clean the cabinet with a mild household cleaner and a non-abrasive cloth or sponge. Do not spill liquid inside the cabinet.

Cleaning or Replacing the Filters (3 Types)

Clean or replace the filters as often as specified in the following paragraphs in order to protect the compressor and extend the concentrator’s life. Replacement filters are available from your Graham-Field equipment provider.

Bottom filter

The bottom filter does not typically require cleaning, but it may if operated in extremely dirty conditions. Perform a visual inspection of the bottom filter every 2 weeks; if the filter is dirty, clean it. The bottom filter is located at the bottom rear of the cabinet (see Figure 1). To clean:

1. Remove the bottom filter

from the squares carefully with your fingers.

2. Inspect the filter; if any part of it is cracked or broken, replace it instead of clean- ing it.

3.If cleaning, rinse filter well

with clear water until all dirt is removed.

4.Dry filter by squeezing between two absorbent towels. Wait for filter to dry. When dry, install filter carefully as shown in Figure 1.

sDamage: Do not operate the concentrator without the filters installed, or while filters are wet. These actions could permanently damage the concentrator.

John Bunn O2 Premier Oxygen Concentrator

13

Page 13
Image 13
Graham Field JB0160-010 Maintenance, Cleaning the Cabinet, Cleaning or Replacing the Filters 3 Types, Bottom filter