8 Specifications
•Flow Range: 1 to 5L/min
•Oxygen Concentration:
•Maximum Outlet Pressure: 6.52 psi ± .065 psi (45kPa ± 4.5kPa)
•Pressure Relief Mechanism Operational at:
36.26 psi ± 3.63 psi (250kPa ± 25kPa)
•Altitude: Up to 5,997 feet (1,828 meters) above sea level without degradation of concentration levels. Efficiency of 90% from 5,997 to 13,123 feet (1,828 to 4,000 meters) above sea level.
•Sound Level: ≤ 55dB
•Power Supply:
•Input Power: 400 Watts Average
•Weight: 57lbs (26kg)
•Dimensions: H 27” x W 18” x D 15”
•Minimum
•Electrical Classification: II Class B Type
•System: Continuous
•
vice): Yellow fault light illuminates when O2 concen- tration is greater than 71% and less than 82%; system
works. Red warning light illuminates when O2 concentra- tion is less than 71%; system shuts off.
Normal Operation Conditions
•Temperature range: 41°F to 104°F (5°C to 40°C)
•Relative humidity: 20% to 60%
•Atmospheric pressure: 12.47 psi to 15.37 psi (86kPa to 106kPa)
John Bunn O2 Premier Oxygen Concentrator | 19 |