Graham Field JB0160-015B-220, JB0160-015-220, JB0160-010-220, JB0160-010B Specifications

Models: JB0160-015 JB0160-015B JB0160-010B

1 28
Download 28 pages 8.54 Kb
Page 19
Image 19

8 Specifications

Flow Range: 1 to 5L/min

Oxygen Concentration: 1-4 LPM 94% ± 2% 5 LPM 92% ± 2%

Maximum Outlet Pressure: 6.52 psi ± .065 psi (45kPa ± 4.5kPa)

Pressure Relief Mechanism Operational at:

36.26 psi ± 3.63 psi (250kPa ± 25kPa)

Altitude: Up to 5,997 feet (1,828 meters) above sea level without degradation of concentration levels. Efficiency of 90% from 5,997 to 13,123 feet (1,828 to 4,000 meters) above sea level.

Sound Level: ≤ 55dB

Power Supply: JB0160-010B, JB0160-010, JB0160-015B,

JB0160-015: AC120V±10% (60Hz) (4.6amps).

JB0160-010B-220, JB0160-010-220, JB0160-015B-220, JB0160-015-220: AC220V±10% (50Hz) (2.2amps)

Input Power: 400 Watts Average

Weight: 57lbs (26kg)

Dimensions: H 27” x W 18” x D 15”

Minimum Warm-Up Time Before Use: 30 minutes

Electrical Classification: II Class B Type

System: Continuous

JB0160-015B, JB0160-015, JB0160-015B-220, JB0160-015-220 only: O2 sensor (electro-chemical de-

vice): Yellow fault light illuminates when O2 concen- tration is greater than 71% and less than 82%; system

works. Red warning light illuminates when O2 concentra- tion is less than 71%; system shuts off.

Normal Operation Conditions

Temperature range: 41°F to 104°F (5°C to 40°C)

Relative humidity: 20% to 60%

Atmospheric pressure: 12.47 psi to 15.37 psi (86kPa to 106kPa)

John Bunn O2 Premier Oxygen Concentrator

19

Page 19
Image 19
Graham Field JB0160-015B-220, JB0160-015-220, JB0160-010-220, JB0160-010B-220 user manual Specifications