TROUBLESHOOTING

Warning: If you experience a problem with your lift and are unable to service it yourself, contact your Lumex dealer.

LF1030 HYDRAULIC PATIENT LIFT

SYMPTOM

POSSIBLE CAUSE

ACTION

Patient lift does not

Air in the cylinder

Bleed the cylinder by turning the release knob

go up or down

 

and push down on the mast. Pump up the lift

 

Loss of hydraulic fluid

again and turn the release knob again.

 

(leaking cylinder)

 

 

 

Replace any leaking cylinder

Patient lift does not

Air in the cylinder

See above Directions

go down

 

 

Note: If the hydraulic cylinder begins leaking fluid, do not operate. Contact your Lumex dealer for a new cylinder.

LF1040 ELECTRIC PATIENT LIFT

SYMPTOM

POSSIBLE CAUSE

ACTION

Control box is not

Battery discharged

Charge battery

responding and relay

Battery charger defective

Replace battery charger

is not clicking

Battery defective

Replace battery

 

Hand control defective

Replace hand control

Actuator does not run,

Actuator plug not properly

Push actuator plug firmly into control box

but relay is clicking

seated

 

 

Actuator defective

Replace actuator

 

Control box defective

Replace control box

Actuator goes up or

Hand control defective

Replace hand control

down, but not both

Control box defective

Replace control box