Haier DW-150W200 After-salesService, Emergency measures, Common troubles and solutions

Models: DW-150W200

1 24
Download 24 pages 34 Kb
Page 16
Image 16

After-sales Service

Analysis on FAQ (frequently asked questions):

If the condenser is dirty or clogged, it could not cool the compressed gas produced by the work of refrigerant compressor. When it is overheated, the compressor may be damaged. Therefore, the condenser must be cleaned on a regular basis to keep clean.

If the freezer is installed close to a wall, it may lead to bad ventilation. In this case, once ambient temperature reaches 32, the condenser temperature would rise rapidly, seriously affecting the refrigeration system. Therefore, such measures must be taken as reinstallation, ventilation improvement or installation of air-conditioning system to ensure that ambient temperature is kept below 27(especially in hot summer or tropical zone).

Unstable voltage or under-voltage may result in abnormal running of compressor. The rated voltage of compressor is 380V±10% (or 220V±10%), beyond which it is recommended to install a voltage stabilizer.

If the freezer is installed in a closed room with air-conditioning system, it may be damaged due to a steep rise of ambient temperature in case of air-conditioner failure, especially on festivals and holidays or at night, so a monitoring system should be installed.

Emergency measures

In case of any fault or any other emergency,

1.immediately move the samples stored in the freezer into a safe freezer in the vicinity.

2.turn off power switch.

3.contact and report the fault to local Haier After-Sales Service Center to have it clear soon.

Common troubles and solutions

The freezer does not work

At first, check whether the air switch is tripped and whether power failure of external power network occurs, if the fault is not caused by the two reasons, check whether the AC contactor coil of compressor control board is energized; if energized, there is nothing wrong with the control board, and then check the circuit to compressor or compressor condition; if not, there is something wrong with the control board.

16-24

Page 16
Image 16
Haier DW-150W200 operation manual After-salesService, Emergency measures, Analysis on FAQ frequently asked questions