Technical Specifications:

Measuring Range:

 

Model 7605:

93.2°F – 109.4°F (34°C – 43°C)

Measuring Accuracy:

+/- 0.4°F (0.2°C): From 96.8°F

 

to 102.2°F (36°C – 39°C)

 

+/- 0.5°F (0.3C): Less than

 

96.8°F (36°C)

 

Greater than 102.2°F (39°C)

Display Resolution:

0.1 (°F or °C)

Response Time:

0.1 Seconds (for temperature reading)

Memory Function:

Displays the last temperature taken

Battery Life:

5000 measurements

Operating Environment:

60.8°F – 104.0°F (16°C – 40°C)

 

with relative humidity up to 95%

 

(non-condensing)

Storage Environment:

-4°F – 122°F (-20°C – 50°C) with

 

relative humidity up to 95%

 

(non-condensing)

Applied Standards

This product conforms to the provisions of the EC directive MDD (93/42/EEC). The following standards apply to design and/or manufacturing of the product:

ASTME 1965-98

Standard Specification for infrared thermometers for intermittent determination of patient temperature

EN 980: 1997

Graphic symbol for use in the labeling of medical devices

EN 60601-1

Medical electrical equipment-Part 1: General requirement for safety.

EN 60601-1-2

Medical electrical equipment-Part 2: Collateral Standard: Electromagnetic compatibility – Requirements and tests.