Page 5
Image 5

Specifications

 

 

 

High Flow

 

Low Flow

Model

 

PM5580: 80/20 heliox

 

PM5480: 80/20 heliox

 

 

PM5570: 70/30 heliox

 

PM5470: 70/30 heliox

Primary Outlet

 

 

15 - 120 lpm

 

3 - 30 lpm

Flow Range

 

 

 

 

 

 

 

 

 

 

 

With both supply pressures at 50 psi (3.4 bar) with

 

 

 

 

 

 

BLEED Closed

 

 

 

Auxiliary Outlet

 

 

2 - 100 lpm

 

0 - 30 lpm

Flow Range

 

 

 

 

 

 

 

 

 

 

 

With both supply pressures at 50 psi (3.4 bar) with

 

 

 

 

 

 

BLEED Open

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Bleed Flow @

 

 

13 lpm or less

 

3 lpm or less

100% FIO2

 

at 50 psi (3.4 bar)

 

at 50 psi (3.4 bar)

Maximum

 

 

 

 

 

 

 

 

 

 

 

Combined

 

 

120 lpm

 

30 lpm

Flow (All Outlets)

 

 

 

 

 

 

 

 

 

 

 

Bypass Flow

 

 

 

 

 

 

 

 

 

 

 

(Loss of Helium or

 

 

> 85 lpm

 

> 45 lpm

Oxygen supply)

 

 

 

 

 

 

 

 

 

 

 

Bypass Alarm

 

50 psi

 

 

60 psi

 

50 psi

 

60 psi

Activation

 

(3.45 bar)

 

(4.14 bar)

(3.45 bar)

 

(4.14 bar)

 

 

13-25 psi

 

16-24 psi

 

18-22 psi

 

16-24 psi

 

 

 

 

 

 

 

 

 

 

 

0.9-1.7 bar

1.1-1.65 bar

1.2-1.5 bar

1.1-1.65 bar

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Alarm Reset:

 

When pressure differential is 6 psi (0.4 bar)

 

 

 

or less.

 

 

 

 

 

 

 

Alarm Sound Level:

 

to 80 db at 1 ft (0.3 m)

 

 

 

Oxygen Concentration

PM5480

 

20 -

 

PM5470

 

30 - 100%

 

Adjustment Range:

 

PM5580

 

100%

 

PM5570

 

 

Gas Supply Pressure:

30 - 75 psi (2.1 - 5.2 bar) Heliox & Oxygen

 

 

 

 

within 10 psi (0.69 bar) of each other

Mixed Gas Stability:

 

±1% Oxygen

 

 

 

 

 

Connection Types:

 

DISS Type - Heliox & Oxygen Inlets & Outlets

 

 

 

and / or NIST Type - Heliox & Oxygen Inlets

 

 

 

 

 

 

 

 

 

 

 

 

Note: All flow-rate values are as measured from an Oxygen flowmeter (uncorrected).

 

 

HELIO2

 

 

Blender

 

 

 

 

4

HELIUM-Oxygen

Page 5
Image 5
Helio PM5400, PM5500 user manual Specifications