2.3.7
2.3.7.11DATA ENTRY Window
Use for drawing texts and graphics on the scanning image.
Fig.
Refer to <3.9.8 Data Entry Function>.
2.3.7.12TILT COMPENSATION Window
Specimen tilting causes defocusing and magnification change in the tilt direction. Tilt compensation is the tool to correct the two phenomena.
Dynamic focus check box
Dynamic focus display/entry box
Dynamic focus micro adjustment button
Dynamic focus adjustment slider
Fig.
(1)Dynamic focus
Checking the Dyn.Focus check box enables dynamic focusing (the function that varies the focusing in tandem with the scanning process). The adjustment slider and the micro- adjustment button can be used to make adjustments so that the image will be in focus over the entire screen.
The possible range of adjustment is ±100(±70°) so that compensation can be made in both positive and negative directions of tilt. These numbers are index values, not angles. Adjustments should be made by viewing the focus on the image.
2 - 47