2.3.7.12TILT COMPENSATION Window

Specimen tilting causes defocusing and magnification change in the tilt direction. Tilt compensation is the tool to correct the two phenomena.

Dynamic focus check box

Dynamic focus display/entry box

Dynamic focus micro adjustment button

Dynamic focus adjustment slider

Tilt-magnification compensation check box

Tilt-magnification compensation display/entry box

Tilt-magnification compensation micro adjustment button

Tilt-magnification compensation adjust slider

Fig. 2.3-36 TILT COMPENSATION Window

(1)Dynamic focus

Checking the Dyn.Focus check box enables dynamic focusing (the function that varies the focusing in tandem with the scanning process). The adjustment slider and the micro- adjustment button can be used to make adjustments so that the image will be in focus over the entire screen.

The possible range of adjustment is ±100(±70°) so that compensation can be made in both positive and negative directions of tilt. These numbers are index values, not angles. Adjustments should be made by viewing the focus on the image.