2.3Graphical User Interface (GUI)

(2)Tilt compensation

Tilting the specimen causes the specimen to be viewed from a tilted direction such that the specimen appears shrunk in the direction of tilt. The tilt compensation function corrects this factor so that the magnification in the direction of tilt will be equal to the situation where there is no tilt.

Checking the Tilt Comp. check box enables this function. The adjustment slider and the micro-adjustment button can be used to set the value in the display/entry box equal to the specimen tilt angle.

See <3.9.9 Raster Rotation, Dynamic Focus and Tilt Compensation>.

2.3.7.13Other

Detail button

Clicking this button opens the Optics tab of the Setup menu.

Show Histogram button

This button opens the Histogram menu.

Scroll Bar and Area Marker Window

Fig. 2.3-37 Scrollbar/Marker Show/Select Window

(1)Show Scrollbar check box

Scrollbars for stigma, brightness and contrast are shown around the scanning image when the Show Scrollbar is checked.

(2)Show Area Marker check box

The area marker, which is a crosshair cursor appearing at the center of scanning image, is shown when the Show Area Marker is checked.

2 - 48