DISPLAY EXPLANATIONS

Date/Time

Indicator

Systolic

Pressure

Diastolic

Pressure

Pulse Rate

Display Symbols:

User 1: Appears when the monitor is operated by User 1.

User 2: Appears when the monitor is operated by User 2.

Weak Battery Symbol: Appears when batteries should be replaced.

Pulse Symbol: Shows the pulse rate per minute.

Memory Symbol: Shows the memory reading number.

Memory Average: Displays average of last 3 readings

WHO Health Indicator: Compares blood pressure against guidelines set by the World Health Organization. See page 21 for more information.

If any of the following letters and numbers appear in the area that diastolic pressure should be displayed, an error has occurred with your reading. See Troubleshooting section of this manual for more information.

Measurement Error: Measure again. Wrap the cuff correctly and keep arm steady during measurement.

Air Circuit Abnormality: Check cuff connection. Measure again.

Pressure Exceeding 300 mmHg: Switch the unit off to clear, then measure again.

Data Error: Remove and reload the batteries.