HoMedics TT-200 Troubleshooting/Error Codes, Condition, Corrective Action, Product Specifications

Models: TT-201

1 16
Download 16 pages 15.37 Kb
Page 8
Image 8
TROUBLESHOOTING/ERROR CODES

TROUBLESHOOTING/ERROR CODES

Condition

Possible Cause

Corrective Action

 

 

 

Consistent low

Temperature

Product Design-

temperature

reading is below

Measurement range is

readings (“Lo”)

measurement range

50˚F (10°C) to 122˚F

 

(below 50˚F/10°C).

(50°C). Contact your

 

 

physician if you have or

 

 

suspect any medical

 

 

problems.

 

The probe is not

Make sure probe is

 

positioned properly

properly positioned

 

in ear.

against the opening

 

 

of the ear canal

 

The waterproof

Clean the probe with a

 

probe

cotton swab moistened

 

is dirty.

with alcohol.

 

 

 

Consistent high

Temperature

Product Design-

temperature

reading is above

Measurement range is

readings (“Hi”)

measurement range

50˚F (10°C) to 122˚F

 

(above 122˚F/50°C).

(50°C). Contact your

 

 

physician if you have or

 

 

suspect any medical

 

 

problems.

 

 

 

Consistent error

Operating

Product Design-

messages (“Err”)

temperature

Operating temperature is

 

is not within the

60.8˚F (16°C) to 104˚F

 

permissible range

(40°C).

 

for measurement.

 

 

 

 

Low battery warning

Battery power

Replace battery (see

 

is low.

“Inserting and Replacing

 

 

Batteries”).

 

 

 

Low battery warning

Battery power is

Replace battery

with “Lo” symbol

too low for the

immediately (see “Inserting

 

thermometer to

and Replacing Batteries”).

 

function.

 

 

 

 

PRODUCT SPECIFICATIONS

Model Numbers

Model Numbers TT-200/TT-201

Power Source

DC 3 Volt (lithium battery CR2032)

 

 

Battery Life

1,000 Measurements in day mode or

300 measurements in night mode.

 

 

 

Sensor

Thermopile

 

 

Temperature Indication

In 4 digits, every 0.1˚F or ˚C

 

 

 

+/- 0.4˚F/0.2°C between 96.8˚F -

 

102.2˚F (36.0°C - 39.0°C)

Accuracy

+/- 0.5˚F/0.3°C less than 96.8˚F

 

(36.0°C) or greater than 102.2˚F

 

(39.0°C)

 

 

Measurement Range

50˚F - 122˚F (10˚C - 50˚C)

 

 

Weight

0.08 lbs (with battery)

 

 

Operating Temperature

60.8˚F - 104˚F (16˚C - 40˚C)

 

 

Operating Humidity

Relative humidity up to 95%

(non-condensing)

 

 

 

Storage Temperature

-4˚F - 122˚F (-20˚C - 50˚C)

 

 

Storage Humidity

Relative humidity up to 95%

(non-condensing)

 

 

 

Contents

Ear Thermometer & Battery, Storage Case,

Instruction Manual

 

 

 

NOTE: These specifications are subject to change without notice. This ear thermometer meets requirements established in ASTM Standard E1965-98.

13

14

Page 8
Image 8
HoMedics TT-200, TT-201 Troubleshooting/Error Codes, Condition, Corrective Action, Product Specifications, Model Numbers