Electrical Connections

50 FT. MAX.

WIRE. 20 FT. OR 50 FT. CABLE LENGTH.

Direct Cell to Analyzer Installation

50 FT. MAXIMUM

49054800SERIESCONDUCTIVITYCELL

Cell to Analyzer through Junction Box

NOTES:

1.FOR PURE WATER SAMPLES IN NON-CONDUCTIVE (PLASTIC, GLASS, ETC) PIPING, GROUND THE BLACK CELL ELECTRODE LEAD NEAR THE CELL. ALTERNATELY, CONNECT TO THE UDA GROUND SCREW AS SHOWN DOTTED. DO NOT GROUND 10, 25, OR 50 CONSTANT CELLS

2.FOR CELL LEADS BLACK AND WHITE, USE 16 TO 22 AWG CABLE SHIELDED TWISTED PAIR, WITH 30pF MAX. CAPACITANCE BETWEEN CONDUCTORS, CONNECT SHIELD TO TERMINAL “10”

3.FOR COMPENSATOR LEADS RED AND GREEN, USE 16 TO 22 AWG, TWO CONDUCTOR CABLE

4.CELL TO ANALYZER CABLES ARE CONSIDERED LOW LEVEL. RUN SEPARATE FROM HIGH LEVEL WIRING.

Figure 4-2 Installation Diagram, 4905 Cells, with 20′ leads directly connected to

UDA2182 Analyzer or connected to Junction Box

12/07

4905 Series Conductivity Cells – Installation and Maintenance

11

Page 17
Image 17
Honeywell 70-82-25-18 manual Direct Cell to Analyzer Installation