INPUT:
l15Vac ±10%, single phase,
OUTPUT:
Two independent outputs, each of which can be set at either
LOAD REGULATION:
Less than 0,01% plus 4mV for a full load to no load change in output current.
LINE REGULATION:
Less than 0.01% plus 4mV for any line voltage change within the input rating.
RIPPLE AND NOISE:
Less than 200µVrms 1mV
TEMPERATURE RANGES:
operating: 0 to 50°C. Storage:
TEMPERATURE COEFFICIENT:
Less than 0.02% plus lmV per degree Centi- grade.
STABILITY.
Less than 0.10% plus 5mV total drift for 8 hours after an initial
INTERNAL IMPEDANCE AS A CONSTANT VOLTAGE SOURCE:
Less than 0.02 ohms from dc to lkHz. Less than 0.5 ohms from lkHz to 1OOkHz. Less than 3.0 ohms from 1OOkHz to lMHz.
TRANSIENT RECOVERY TIME:
Less than 50µsec for output recovery to within 10mV following a full load current change in the output.
OVERLOAD PROTECTION:
A fixed current limiting circuit protects the power supply for all overloads including a direct short placed across the terminals in constant voltage operation.
METERS:
Each front panel meter can be used as either a
OUTPUT CONTROLS:
RANGE switches select desired operating mode for each section and coarse and fine VOLTAGE controls set desired output voltages.
OUTPUT TERMINALS:
Six
ERROR SENSING:
Error sensing is normally accomplished at the front terminals if the load is attached to the front or at the rear terminals if the load is attached to the rear terminals. Also, provisions are included on the rear termina1 strips for remote sensing.
REMOTE RESISTANCE PROGRAMMING: 200 ohms per Volt.
REMOTE VOLTAGE PROGRAM MING: 1 Volt per Volt.
COOLING:
Convection cooling is employed. The supply has no moving parts.
SIZE:
3 ~ ½ " H x
can be mounted side by side in a standard 19” relay rack.
WEIGHT:
10 lbs, net, 13 lbs. shipping.
FINISH:
Light gray front panel with dark gray case.
POWER CORD:
A