| ERROR CODES |
| |||
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
ERROR | DESCRIPTION |
| PROBABLE CAUSE |
| |||
CODE |
|
| |||||
|
|
|
|
|
|
| |
|
|
|
|
|
|
|
|
Er 6 | SHORTED THER- |
| Occurs if the thermocouple wires are shorted or the heater plates are |
|
|
| |
| MOCOUPLE OR |
| defective causing Infinity to achieve a HEAT RATE of less than 6% of the |
|
|
| |
| DEFECTIVE HEATER |
| maximum attainable HEAT RATE with full power applied to the heater |
|
|
| |
| PLATES |
| plates for 5 minutes. This error will also occur if a program is running |
|
|
| |
|
|
| and the door is kept open for more than 5 minutes or the triac or the |
|
|
| |
|
|
| Main PC board developed a problem during a burnout process. |
|
|
| |
|
|
|
|
|
|
| |
|
|
|
|
|
|
|
|
|
|
|
|
|
| ||
Er 7 | THERMAL |
| Occurs when the temperature has exceeded 2015°F (1102°C) for 1 |
| |||
| RUNAWAY |
| minute if the highest temperature programmed was less than 2000°F |
|
| ||
|
|
| (1093°C). This also occurs when the temperature has exceeded |
|
| ||
|
|
| 2050°F (1121°C) for 1 minute if the highest temperature programmed |
|
| ||
|
|
| was 2000°F (1093°C) or higher. |
|
| ||
|
|
|
|
|
|
| |
|
|
|
|
|
|
|
|
Flashing | INVALID ENTRY |
| Occurs when the HEAT RATE in STAGE 2, 3 or 4 is set to “COOL” but |
|
|
| |
99:99 | ERROR WHILE A |
| the corresponding TEMP is entered to heat to a higher temperature |
|
|
| |
when | PROGRAM IS |
| than the previous stage. |
|
|
| |
NIGHT | RUNNING |
|
|
|
|
|
|
|
|
|
|
|
| ||
TIME |
|
|
|
|
|
| |
|
|
|
|
|
|
| |
(DELAY |
|
|
|
|
|
|
|
START) is |
|
|
|
|
|
|
|
pressed |
|
|
|
|
|
|
|
during a |
|
|
|
|
|
|
|
program |
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
18