I-Transia™

3.Lower the vertical poles to the lowest position. Refer to 3.6 Adjusting the Height, page 5 .

4.Reverse the assembly steps to disassemble the system. Refer to 3.3 Assembling the Freestanding Rail, page 3 .

It may be necessary to push the top bracket upwards in order to separate it from the vertical pole module.

4.2 Transport

1.Disassemble the freestanding rail system. Refer to 4.1 Disassembly, page 5 .

2.Put the components of the system into the portable carrier bag (optional accessory).

4.3 Storage

CAUTION!

Risk of Product Damage

The freestanding rail system should be stored in a room where the humidity does not exceed 70%.

Do not store the system in a bathroom.

5 Maintenance

5.1 Cleaning

CAUTION!

Risk of Product Damage

DO NOT autoclave or use harsh chemicals when cleaning the system.

Clean the freestanding system with a damp cloth using ordinary household detergents and household disinfectant.

To prevent cross-infection, the parts of the hoist that come in contact with caregivers and patients must be cleaned and disinfected. A soft cloth, dampened with water and a small amount of dish liquid, is all that is needed to clean the hoist, lifting strap and carabiner. The hoist can be cleaned with non-abrasive cleaners.

6 Troubleshooting

6.1 Troubleshooting Table

PROBLEM

SOLUTION

PROCEDURE

 

 

 

The system is not

Check that the

Refer to 3.3

stable.

feet are mounted

Assembling the

 

correctly.

Freestanding Rail,

 

 

page 3 .

 

Check the vertical

 

 

 

poles are pushed

 

 

securely into the top

 

 

bracket.

 

 

 

 

 

Check that the rails

 

 

are level.

 

 

 

 

 

Check that the

 

 

top brackets are

 

 

tightened.

 

 

 

 

 

Check that the top

 

 

brackets have been

 

 

positioned correctly.

 

 

 

 

 

Check that the

 

 

yellow thumb

 

 

screws are

 

 

tightened.

 

 

 

 

 

Check the top brackets for defects or

 

deformities. If found, contact an Invacare

 

dealer or certified technician.

 

 

 

The height

Using the 5 mm

Refer to 3.6

adjustment is not

Allen key, check that

Adjusting the Height,

working.

the vertical pole is

page 5 .

 

not stuck in the far

 

 

end position.

 

 

 

 

 

If the top part of the vertical pole is not

 

moving when turning the Allen key, stop

 

using the system and contact an Invacare

 

dealer or certified technician.

 

 

 

7 Technical Data

7.1 Typical Product Parameters Freestanding rail system

Minimum Dimensions

2020 x 2720 mm (79.5 in x 107 in)

(HxW)

 

Maximum Dimensions

2820 x 3890 mm (111 in x 153 in)

(HxW)

 

Maximum lifting

250 kg (550 lb)

capacity

 

Total weight without hoist

Freestanding rail

52.0

kg (114.5 lb)

system

 

 

 

Vertical pole, each

13.5

kg (30

lb)

Top brackets and rail

25.0

kg (55

lb)

Materials

Rails and posts

Anodized aluminum

Feet

Powder varnished steel

Brackets

Powder varnished steel

6

1183379A

Page 6
Image 6
Invacare 1183379-A user manual Maintenance, Troubleshooting, Technical Data