USING THE SCALE
Part No. 1150696 23 Traverse Lift/Transporter
Calibrating the 1100 Bath Scale
NOTE:Forthisprocedure,refertoFIGURE3.1onpage20andFIGURE 3.4.
1. ThepatientMUSTberemovedfromthebathlifttoproperlycalibratethebathscale.Refer
toTransferringinto/outoftheLiftoftheBathLiftOwner’sManualforinstructions.
2. Removetheseat/stretcherfromthebathlift.
•Lockthetworearlockingcasters.
•Pressonthedownarrowonthehandpendanttolowerthebathlift.
•Removethethreeboltsthatsecurethehangerrodstothebathscale.
•Pullthehangerrodsawayfromthepatientscaletoremovetheseatorstretcherfrom
thebathlift.
3. Removethefourscrewsthatattachthecovertothebathscale.
NOTE:TheRUN/CALswitchislocatednexttothedisplaywindowandisvisibleafterthecoveris
removed.
4. SlidetheswitchmarkedRUN/CALtotheCALposition.
5. AdjustR9sothedisplayindicates+15poundoffset.RefertoDetail“A” ofFIGURE3.4.
6. ReverseSTEP2toattachtheseatorstretchertothebathlift.
7. Makenoteoftheweightoftheseatorstretcheronthedisplay.
8. Put200poundsofcalibratedweightontheseatorstretcher.Checkfortheproperweight
increaserelativetotheweightoftheseatorstretchernotedinSTEP7.
9. AdjustR18untiltheweightonthedisplayisaccurateofthecombinedweightofthe
calibratedweightandtheweightoftheseatorstretcher.RefertoDetail“A” ofFIGURE3.4.
10. Removethecalibratedweightfromthebathscale.RepeatSTEPS59asnecessary.
11. SlidetheRUN/CALswitchtotheRUNposition.PresstheZERObutton.
12. Whenthedisplaywindowreads0.0,placethecalibratedweightontheseatorstretcher
andmakesuretheweightonthedisplayiscorrect.
13. Removethecalibratedweightfromthebathscale.
NOTE:Thebathscaleisreadyforuse.
FIGURE 3.4 Calibrating the 1100 Bath Scale
Screw
(1 of 4)
Cover
R9 Adjustment Screw
R18 Adjustment Screw
DETAIL “A” - 1100 BATH SCALE CIRCUIT BOARD