SECTION 4—INSPECTION AND COMPRESSOR INTERFACE

External Examination

1.Examine the outside of the cylinder for the following conditions and replace the cylinder if they exist:

Dents or dings

Arc burns

Oil or grease

Any other signs of damage that might cause a cylinder to be unacceptable or unsafe for use.

2.Examine the cylinder for evidence of fire or thermal damage. Evidence includes charring or blistering of the paint, or other protective coating or heat sensitive indicator. If fire or thermal damage is found, replace the cylinder.

3.Inspect the cylinder fillport for the following and replace if found:

Debris, oil or grease

Noticeable signs of damage

Signs of corrosion inside the valve

Signs of excessive heat or fire damage

Part No 1163156