SECTION 4—INSPECTION AND COMPRESSOR INTERFACE

Filling the Cylinder

NOTE: For this procedure, refer to FIGURE 4.1 thru FIGURE 4.6.

1.Check the cylinder pressure before attempting to fill cylinder. Refer to Checking Cylinder Pressure on page 12.

2.Close the cylinder valve by rotating clockwise (FIGURE 4.1).

3.Remove the conserver or regulator according to the device manufacturer instructions (FIGURE 4.1).

CONSERVER/REGULATOR

INSTALLED

Conserver/

Cylinder Regulator

Valve

CONSERVER/REGULATOR

REMOVED

Gauge

Cylinder

FIGURE 4.1 Component Identification

4.Remove the post valve cover (located on the cylinder) and the connector fillport covers (located on the compressor) (FIGURE 4.2).

Part No 1163156

13

HomeFill® Cylinder