SECTION 4—INSPECTION AND COMPRESSOR INTERFACE

NOTE: If the outer ring (sleeve) is in the UP position (GREEN dots not visible), the connector fillport will not be able to accept the cylinder fillport. Pushing DOWN momentarily will reset the connector fillport (GREEN dots visible) to accept the cylinder fillport.

PUSH

Outer Ring (Sleeve)

DOWN

Connector Fillport

GREEN Dots

 

BEFORE

COUPLING

CYLINDER,

PUSH DOWN

ON SLEEVE

UNTIL

GREEN DOTS

ARE VISIBLE.

FIGURE 4.3 Resetting Connector Fillport

6.Grasp the cylinder assembly as shown in Detail “A” in FIGURE 4.4.

7.Position the cylinder in the compressor cradle as shown in Detail “C” in FIGURE 4.4.

8.Align the cylinder fillport with the connector fillport as shown in Detail “C” in FIGURE 4.4.

9.Pull UP on the outer ring (sleeve) of the connector fillport (Detail “B”) while pushing DOWN on the cylinder assembly to couple the cylinder fillport into the connector fillport as shown in Detail “B” in FIGURE 4.4.

NOTE: The cylinder is properly connected when an audible “click” is heard.

10. Press the compressor power switch to the On ( ) position.