SECTION 6—TYPICAL PRODUCT PARAMETERS

Oxygen Cylinder Duration

The total delivery of oxygen via the conserver is related to the breathing rates, it is user adaptive in that the total oxygen delivered per minute will automatically adjust with user need, as expressed by increased or decreased breathing rates. For example, at all settings, twice as much oxygen per minute will be delivered if one breathes twenty (20) times per minute as compared with ten (10) times per minute. Refer to OXYGEN DURATION TABLE on page 25.

OXYGEN DURATION TABLE

 

SETTING

1

2

3

4

5

6

Continuous

 

 

 

 

 

 

 

 

Flow

 

 

 

 

 

 

 

 

2 L/min.

 

 

 

 

 

 

 

 

 

Cylinder Type

Cylinder

 

 

 

 

 

 

 

 

Volume

 

 

 

 

 

 

 

 

Liters @

 

Estimated Cylinder Duration in Hours

 

2015 psi

 

 

(Based on 20 breaths/min.)

 

 

 

 

 

 

 

 

 

M2

36

1.9

.9

.6

.5

.4

.3

.3

 

 

 

 

 

 

 

 

 

M4(A)

113

5.9

2.9

2.0

1.5

1.2

1.0

.9

 

 

 

 

 

 

 

 

 

M6(B)

164

8.5

4.3

2.8

2.1

1.7

1.4

1.4

 

 

 

 

 

 

 

 

 

ML6

171

8.9

4.5

3.0

2.2

1.8

1.5

1.4

 

 

 

 

 

 

 

 

 

M9(C)

246

12.8

6.4

4.3

3.2

2.6

2.1

2.1

 

 

 

 

 

 

 

 

 

D

425

22.1

11.1

7.4

5.5

4.4

3.7

3.5

 

 

 

 

 

 

 

 

 

E

680

35.4

17.7

11.8

8.9

7.1

5.9

5.7

 

 

 

 

 

 

 

 

 

Part No 1160840

25

Pneumatic Oxygen Conserver

Page 25
Image 25
Invacare IOC100P manual Oxygen Cylinder Duration, Oxygen Duration Table