SECTION 6—TYPICAL PRODUCT PARAMETERS

SECTION 6—TYPICAL PRODUCT PARAMETERS

OXYGEN DELIVERY:

SELECTOR

 

LITER

 

SWITCH

 

FLOW

 

POSITION

 

EQUIVALENCY

 

 

 

 

 

0

 

OFF

 

 

 

 

 

1

 

1

 

 

 

 

 

2

 

2

 

 

 

 

 

3

 

3

 

 

 

 

 

4

 

4

 

 

 

 

 

5

 

5

 

 

 

 

 

6

 

6

 

 

 

 

CONTINUOUS FLOW

 

 

 

EMERGENCY BYPASS

 

 

 

SETTING:

Factory Preset at 2 ± .5 L/min.

 

 

 

 

REQUIRED OPERATING

 

 

 

PRESSURE:

200 psi to 3000 psi (13.8 bar to 206.8 bar)

 

 

REGULATOR:

Built-in, 25 ± 5 psi (1.7 bar ± .34 bar), brass

 

high pressure with aluminum low

 

pressure materials

 

 

 

 

DIMENSIONS:

 

 

 

 

 

Weight

Approximately 14.8 ounces (420 grams)

 

 

Overall Length

Approximately

 

5” L (12.7 cm) x 2.5” H (6.4 cm) x 2.5” W

 

 

(6.4 cm)

 

 

 

 

OPERATING CONDI-

 

 

 

TIONS TEMPERATURE:

32°F (0°C) to 122°F (50°C)

 

 

 

 

OPERATING RELATIVE

 

 

 

HUMIDITY:

15% to 95%

 

 

OPERATING ALTITUDE:

0 to 10,000 feet (0 to 3,048 meters)

 

 

 

 

Part No 1160840

23

Pneumatic Oxygen Conserver

Page 23
Image 23
Invacare IOC100P manual Typical Product Parameters, Regulator