MAINTENANCE

2.Clean the cabinet filters with a vacuum cleaner or wash in warm soapy water and rinse thoroughly (FIGURE 10).

3.Dry the filters thoroughly before reinstallation.

CAUTION

DO NOT operate the concentrator without the filters installed.

FIGURE 10 - CLEANING THE CABINET FILTERS

CABINET

WARNING

UNPLUG the concentrator when cleaning. To avoid electrical shock, DO NOT remove cabinet.

1.Clean the cabinet with a mild household cleaner and non-abrasive cloth or sponge.

HUMIDIFIER

NOTE: To clean the oxygen humidifier, follow the instructions provided by the manufacturer. If none are provided, follow these steps:

1.Clean the humidifier every day.

2.Wash it in soapy water and rinse it with a solution of ten (10) parts water and one (1) part vinegar.

3.Rinse thoroughly with hot water and refill with distilled water to the level shown on the humidifier.

WARNING

DO NOT overfill Humidifier Bottle.

18

Page 18
Image 18
Invacare IRC5LXO2, IRC5LX manual Maintenance, Humidifier