SPECIFICATIONS

SPECIFICATIONS

Electrical Requirements:

Rated Current Input:

Sound Level:

Altitude:

*Oxygen Output Concentration Levels:

120 VAC +10% (132 VAC/108 VAC),

60Hz

4.0 A

50 dB Average

Up to 6,000 ft. (1828 meters) above sea level without degradation of concentration levels. From 6,000 ft. (1828 meters) to 13,129 ft. (4000 meters) below 90% efficiency.

ALL 5LXO2/5LX MODELS 95.6% to 87% at 1/2 to 5 L/min.

*Stated concentration levels achieved after initial warm-up period (approximately 30 minutes).

Maximum Outlet Pressure: 5 + 0.5 psi.

 

 

 

(34.5 + 3.45 kPa.)

 

SensO2® Alarm Thresholds:

 

 

 

 

 

 

 

 

 

 

Label

O2 Purity

 

Indicators

 

 

Symbol

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Below 73%

 

RED Indicator Light -

 

 

 

 

Continuous Audible Alarm

 

 

 

 

 

 

 

 

 

 

Sieve-GARD™Compressor

 

 

 

 

 

Shutdown

 

 

 

73% to 85%

 

YELLOW Indicator Light

 

 

O2

Over 85%

 

GREEN Indicator Light

 

 

 

 

 

 

 

 

 

 

 

 

 

8

Page 8
Image 8