SECTION 6—OPERATING INSTRUCTIONS

SensO2 Oxygen Purity Indicator - Model IRC5PO2AW

This feature monitors the purity of oxygen generated by the oxygen concentrator. If purity falls below factory preset standards, indicator lights on the control panel will illuminate.

Initial Startup of the Concentrator

NOTE: Concentrator may be used during the initial start warm‐up time (approximately 30 min.) while waiting for the O2 purity to reach maximum.

When the unit is turned on, the GREEN light will come on (SYSTEM OK/O2 greater than 85%). After five minutes, the oxygen sensor will be operating normally and will control the indicator lights depending on oxygen concentration values.

Explanation of Oxygen Purity Indicator Lights - Model IRC5PO2AW

NOTE: For this procedure, refer to FIGURE 6.6 on page 23.

GREEN light (O2) ‐ Normal Operation.

YELLOW light ( ) ‐ Immediately call supplier. You may continue to use the concentrator unless instructed otherwise by your supplier. Be certain that backup oxygen is nearby.

RED light ( ) ‐ Total unit shutdown. Immediately switch to backup oxygen supply and call supplier.

GREEN light ‐ with YELLOW light flashing ‐ Immediately call supplier. Oxygen sensor malfunctioning; you may continue to use the concentrator.