Precautions on installing medical electrical equipment (Safety and hazard prevention)

For the safe use of this equipment, be absolutely sure to follow the guidelines below.

1.Leave installation of the Mobile Video Cart and the unit to a specially trained service man.

2.Be absolutely sure to use the separately sold AA-V31E isolation transformer and AC adapter AA-V112E.

3.There are openings in the equipment. Fire or electric shock may result if liquid splashes inside one of these openings. The front of the JVC-specified video cart has an opening. To prevent any possible prob- lems, please set the cart in a reasonable position and use carefully. JVC cannot guarantee the total system when using an unspecified video cart.

Cart information

Please use the VS5402.X or VS6300.X Mobile Video Cart to house the equipment.

(1)Maximum load

Additional load: 225 kg

Monitor shelf: 35 kg

(2)Front two casters have brakes.

When using mobile video cart, two brakes should be set for safety after setting the position of the cart.

(3)Important notice 1

When using a mobile cart in OR, as the cart has no front cover, precautions to avoid splashing at front area of it should be taken.

When moving a mobile cart in OR, precautions to avoid crossing over the power cord of the cart.

When using a mobile cart in OR, precautions to avoid power cord of the cart to be caught by foot should be taken.

(4)Important notice 2

When mounting monitor on a mobile cart, we strongly recommend the use of a monitor shelf and monitor strap for safety. (Refer to the following figure.)

(5)Equipotential terminals are located inside the cart

The back panel of the cart can be easily opened without any tool to gain access to the terminals.

ex.) Mobile Video Cart VS5402.X

For medical use, please use the JVC Isolation Transformer, AA-V31E, and JVC AC Power Adapter, AA-V112E, in order to meet the following classifications.

(1)Class 1 Equipment powered by an external electrical power source.

(2)Type BF Applied part (Degree of protection against electric shock) The specifications for the AA-V31E and AA-V112E are given on page 26.